|
|
OPERATION OF THE PHILIPS CM-200 TRANSMISSION ELECTRON MICROSCOPE
by
Birute Jakstys and Anna Klintsova (2nd revision, 01/2003). When not in use, the CM-200 should be in the MICROSCOPE ON (see 1. below)
configuration with the HIGH TENSION OFF. The microscope is normally never
turned off.
Please record your name, other requested
specifics, and comments in the logbook.
ITG users may reserve time on the TEM using the TEM calendar.
| Initiating Operation: |
- When the CM-200
is ON, the white STAND BY and red MICROSCOPE OFF buttons will be
illuminated; the ON button will be dark. The lit buttons indicate
their availability as emergency functions while the microscope is
running.
- Depress the
PANEL DIM knob to illuminate the data monitor screen and emission gauge.
Clockwise rotation of this knob increases the intensity of the panel
lighting and the light on the flexible stalk. Clockwise rotation of the
DATA DIM knob increases the brightness of the data monitor screen.
- Ensure that the
UHV and HIVAC indicator lights are lit (green). [See 7. below if the two
indicator lights are not lit.]
- (Anticontamination
Device) Place the metal-clad glass vacuum dewar into its holder to the
right of the column so that the soft copper wire "beard" hangs inside
it. The glass viewing windows should be covered for safety. Fill the
dewar nearly full with liquid nitrogen and place the styrofoam cap on
top.
- The data monitor
screen will display the CM-200 PHILIPS MICROSCOPE STATUS startup page
or, more likely, one of the MODES or MODE SELECTION pages. If the READY
light (on the pushbutton directly below the data monitor screen, to the
right) is lit, pressing it will move the screen selection back in the
hierarchy of pages. Press the MODES key to obtain the MODE SELECTION
page.
- Press the TEM
LOW DOSE key on the MODE SELECTION page. [Unless we are taking film
negatives, we now use the 'scope almost exclusively in TEM LOW DOSE
BRIGHTFIELD mode, with SEARCH selected. This is because the TVIPS
digital imaging software was written to be operated in this mode.] If
the letters TEM LOW DOSE are highlighted it will be necessary to press
only once; otherwise press the TEM LOW DOSE key twice. The screen should
now display the TEM LOW DOSE BRIGHTFIELD page.
- Press the VACUUM
key to check the vacuum status of the instrument. It should read READY
at the top center of the data monitor screen. If it reads START-UP the
operator must wait for the vacuum to improve before the 'scope can be
used. [If you find the microscope stuck at START-UP, e.g., if (on the
VACUUM page) P3 reads less than 63 but IGP is at 0 (in this case
indicating that the ion getter pump is off) and the IGP icon is not
highlighted on the schematic, press VACUUM SYSTEM OFF, wait a second or
two, and then press VACUUM SYSTEM ON.] Do not use the microscope unless
the ion getter pump (IGP) reading is lower than 20.
- To return to the
TEM LOW DOSE BRIGHTFIELD page press the READY button.
- To select the
operating voltage, press the PARAMETERS key to open the PARAMETERS
pages. On the first of these pages the kV may be modified by pressing
the left (lower) or right (higher) key adjacent to the HIGH TENSION kV
notation on the screen. Most TEM users in this laboratory work at 120
kV.
- Also on the
first PARAMETERS page is the EMISSION setting, which should ordinarily
be left at 1. A higher value will produce a brighter beam image but will
also shorten the lanthanum hexaboride (LaB6) filament life.
- Press the READY button to return to
the TEM LOW DOSE BRIGHTFIELD page.
- Press the HIGH
TENSION ON/OFF button (illuminates green when the high tension is on).
- With the high
tension on, an operator may load previously stored or default
alignment/mode/stigmator settings for the kV at which the 'scope is
operating. This requires logging onto the computer (the one with the
keyboard and mouse on the movable rest and the CRT on top of the control
panel), opening TCL from QuickFind, in the Start menu, and accessing the
EM MENU window (F4). At this point, you may select ALIGNMENTS from the
EM MENU window and load previously stored settings.
- IMPORTANT: If
you intend to expose TEM negatives using the built-in camera, the CCD
camera control unit, on top of the instrument panel just to the left of
the column, must be switched to CM rather than CCD. Otherwise the beam
blank will remain on and EXPOSURES WILL NOT BE MADE. Because we are now
using the CCD camera almost exclusively, film exposure/processing is not
covered in this set of instructions.
|
| Beam
Alignment: |
These functions are
performed either without a specimen holder or with the specimen holder
inserted fully into the 'scope but held just out of the electron beam by
wedging a pen or pencil vertically between the specimen holder handle and
the face of the Compustage. If a specimen holder is not in place, a
radiation safety interlock keeps the spot size at 5 or a higher number
(indicating a smaller spot size; the largest and brightest spot size is 1).
The interlock prevents a larger spot size from being selected and produces a
much dimmer beam than that seen with the holder in place. Note: A specimen
holder (without a specimen or with a specimen but wedged just out of the
electron beam) must be in place in order to adjust the C2 (COND) stigmator
(8. in the list below).
- Using the
MAGNIFICATION knob, select 17,500x.
- Press the MODES
key to access the MODE SELECTION page. From there, press CONFIGURATION.
The CONFIGURATION page displays the
filament heating status and provides a schematic of the
available apertures (aperture
memo) for the condensor 2 (C2), objective, and selected area
lenses. Under CATHODE, at the top of the page, LaB6 should be
highlighted, indicating that a LaB6 filament is in use.
filament should be
highlighted next to the number (usually between 22 and 30) indicating
the saturation limit of the filament. With the limit set (highlighted),
the operator can turn the FILAMENT knob clockwise indefinitely without
fear of oversaturating the filament. Turn the filament up to a number
that is 4 or 5 less than the indicated filament limit.
- Defocus the
electron beam by turning the INTENSITY knob clockwise to strongly
overfocus the C2 lens. Bring the beam back to crossover (the smallest
image of the beam) and center it using the SHIFT X/Y knobs
- Choose a C2
aperture (usually position 3, which is a 100-micron aperture) by
rotating the largest knurled knob on the topmost aperture control on the
column to the desired position.
- Using
the INTENSITY knob, slowly overfocus (turn clockwise) the beam to
coincide with the diameter of the intermediate-sized black circle on the
phosphorescent viewing screen. If the defocused beam is not coincident
with the circle, adjust it using the C2 aperture centering controls.
These are both the intermediate-sized knurled knob on the aperture rod
and the knurled knob on the side of the assembly (to the right). DO NOT
TWIST THE SMALLEST, INNERMOST KNOB (WHICH IS NOT KNURLED) BECAUSE IT
UNSCREWS THE APERTURE ROD. Repeat these steps (3 and 5) until the beam
spreads uniformly around the reference circle while being overfocused
(INTENSITY knob).
- Using the
INTENSITY knob, adjust the beam to crossover. Depress the FINE button
(to the left of the INTENSITY knob; the indicator light turns green when
it is on) and use the INTENSITY knob to sharpen the desaturated filament
image. Center the image using the SHIFT X/Y knobs.
- Press the STIG
button, on the right-hand instrument panel, to open the STIGMATOR
CONTROL page.
- Press the COND
key on the data monitor screen to select the C2 lens for astigmatism
correction. Use the MULTIFUNCTION X/Y knobs, while rotating the beam
through crossover using the INTENSITY knob, to obtain the cleanest,
roundest desaturated filament image possible. [If a tungsten instead of
a LaB6 filament was in the 'scope, this is where the
desaturated filament image would appear like a central bright spot with
a halo around it, similar to the CBS TV logo.]
- Press the STIG
button to return to the CONFIGURATION page.
- Turn the
FILAMENT knob clockwise to fully saturate the filament. When the
saturation point (as defined by the FIL LIMIT) has been reached, the
microscope will emit a 'beep.'
- Press the READY
button and then the TEM LOW DOSE key to return to the TEM LOW DOSE
BRIGHTFIELD page.
- Using the lever
to the left of the viewing chamber, lower the small phosphorescent
screen into place. Insert the beam stop (at the right-hand side top of
the viewing chamber) so that it will be visible across the small screen.
Use the binoculars to observe the beam stop. To adjust the binoculars
for your eyes, first adjust the interpupillary distance so that you can
see through them with both eyes; then adjust each eyepiece so that the
rough edges of the beam stop are in focus for each eye. When you are
done, retract the beam stop and lift the small screen back out of view.
|
| Specimen
Holder Removal, Loading, and Insertion: |
Never
remove or insert the specimen holder when the red indicator light (on the
front of the compustage housing) is on. Do not touch the leading edge of the
specimen holder (from the o-ring to the tip) with ungloved hands. This
portion resides inside the vacuum and must be kept clean.
- REMOVAL: To
remove the specimen holder from the column, carefully pull the round
black handle straight out until it stops, and hold it firmly so that it
does not get pulled back into the 'scope by the vacuum. Now rotate it
clockwise until it stops again; it may now be pulled straight out
(carefully), free of the column. The specimen holder should be set down
only on its Lucite stand.
- SPECIMEN LOADING:
- Use the pin tool (located in the Lucite stand, under the tip)
to lift the grid clamping device at the tip of the speciment holder.
-
Transfer your grid to the specimen holder using forceps. To
make scanning the grid easier, you may wish to orient one set of
grid bars parallel to the long axis of the specimen holder.
-
Use the pin tool to carefully lower the clamping device onto the
grid and lock it in place.
- INSERTING THE SPECIMEN HOLDER INTO
THE COMPUSTAGE:
- With the
small pin on its shaft at the 4 o'clock position, carefully insert
the specimen holder into the airlock entryway at the center of the
compustage. Insertion of the holder will initiate the prepumping
sequence, and the red indicator light on the front of the compustage
will come on. Slide the holder in until it stops and feels fairly
solid to the touch; at this point it will not go all the way in.
- The data
monitor screen will automatically open to the HOLDER SELECTION PAGE.
Press the appropriate key (in most cases you'll choose NO COMPUSTAGE B
TILT) and then press the READY button (below the data monitor screen).
The data monitor screen will return to the previously selected page
(normally TEM LOW DOSE BRIGHTFIELD).
After the
red indicator light goes out, grip the specimen holder firmly by its
black plastic handle and rotate it counterclockwise. When it reaches
the limit of its rotation, the vacuum will begin to pull the holder
deeper into the column. Maintain a firm grip on the handle as the
holder enters the column so that it does not get drawn in too quickly.
Once the holder appears to be all the way in, jiggle it very gently to
ensure that it is indeed inserted completely into the column.
|
| Eucentric
Height Adjustment: |
- Set the
magnification to 5,800x
and use the X/Y JOYSTICK to center a small notable feature of your
specimen.
- Focus your
chosen feature using the knobs marked STEP SIZE. The outer knob changes
the focus; the inner knob (the step size adjustment) modifies the amount
of focus change per 'click' of the outer knob.
- From the TEM LOW
DOSE BRIGHTFIELD page, press COMPUSTAGE once; the COMPUSTAGE REGISTER
CONTROL page will appear.
- Press A-WOBBLER;
this will initiate back-and-forth tilting of the goniometer.
- Use the Z
control lever on the JOYSTICK to move the specimen up or down, whichever
is appropriate to minimize apparent movement of the centered feature.
- When the feature moves only
minimally or not at all, press the A-WOBBLER key to inactivate tilting;
then press the READY button to return to TEM LOW DOSE BRIGHTFIELD.
|
| Pivot
Point Alignment: |
Ensure that the specimen is eucentric
before performing this procedure. For this procedure you may work with the
OBJECTIVE APERTURE in place to protect your specimen.
- Center (X/Y
JOYSTICK) and focus (concentric knobs under STEP SIZE) an image feature
at 24,500x (MAG knob).
- Press the ALGN
button to access the ALIGNMENT SELECTION page.
Note
that the alignments are divided into PROCEDURES (on the left side of the page)
and DIRECT alignments (most of which are on the right side of the page). Most
users will not want to access the PROCEDURES, which are long and complicated.
For this set of instructions we will be using only the DIRECT alignments.
- Using the
INTENSITY knob, adjust the beam to crossover.
- Press the
beamcoils PIVOT POINT X key, on the right side of the page, so that it
is highlighted.
- Using the
MULTIFUNCTION X/Y knobs, bring the two beam spots (the pivot points, on
the phosphorescent viewing screen) together so that they overlap.
- Center the
coinciding spots using the SHIFT X/Y knobs.
- Press the
beamcoils PIVOT POINT Y key.
- Using the
MULTIFUNCTION X/Y knobs, bring the two beam spots together so that they
overlap.
- Center the
coinciding spots using the SHIFT X/Y knobs.
- Press the ALGN
button to exit the ALIGNMENT SELECTION page.
|
| Rotation
Center Alignment: |
This
procedure may be performed with the OBJECTIVE APERTURE in place to protect
the specimen and add contrast to the image.
- Focus and center
a feature of the specimen at 100,000x.
- Press ALGN to
open the ALIGNMENT SELECTION page. On the upper right-hand side of the
page, press ROT CENTER so that it becomes highlighted. Now either
voltage or current centering may be performed; it is not
necessary to do both.
VOLTAGE CENTERING:
- On the lower
right-hand side of the page, select VOLT (under rot center VOLT CURR) so
that it becomes highlighted. This will cause the high tension to
modulate (the inner STEP SIZE knob adjusts the amplitude of modulation).
If the chosen feature shifts off center laterally, the beam is not
aligned along the optical axis of the microscope and must be corrected.
- Use the
MULTIFUNCTION X/Y knobs to stabilize the feature at the center of the
screen, eliminating all lateral movement. The feature should appear to
be pulsating.
- Press the ALGN
button to return to the TEM LOW DOSE BRIGHTFIELD page.
CURRENT CENTERING:
- On the lower
right-hand side of the page, select CURR (under rot center VOLT CURR) so
that it becomes highlighted. This will cause the objective lens current
to modulate (the inner STEP SIZE knob adjusts the amplitude of
modulation). If the chosen feature shifts off center laterally, the beam
is not aligned along the optical axis of the microscope and must be
corrected.
- Use the
MULTIFUNCTION X/Y knobs to stabilize the feature at the center of the
screen, eliminating all lateral movement. The feature should appear to
be pulsating.
- Press the ALGN
button to return to the TEM LOW DOSE BRIGHTFIELD page.
|
| Centering
the Objective Lens Aperture: |
For
this procedure a specimen must be in place. Choose an area for which it is
acceptable to sustain beam damage.
- With the
OBJECTIVE APERTURE out, set the magnification to 5,800x and overfocus
the beam by turning the INTENSITY knob clockwise from crossover.
- Set the focus
step size (inner STEP SIZE knob) to 5 and press the D button to put the
'scope in diffraction mode.
- If
necessary, adjust the camera
length to 620 mm using the MAG knob. [Note that you can
accidentally choose 6.2 m rather than 620 mm.]
- Center the
diffraction spot using the MULTIFUNCTION X/Y knobs.
- Using the focus
(outer STEP SIZE) knob, refocus the beam to the smallest, brightest
spot. It may also be necessary to adjust the INTENSITY knob..
- Insert
the OBJECTIVE APERTURE by rotating the aperture displacement lever below
it to the left.
- Apertures may be
selected by rotating the largest knurled knob on the objective aperture
assembly to any one of four numbered positions. (The APERTURE MEMO lists
the diameters and positions of the apertures currently installed in the
'scope. It may be accessed from the TEM LOW DOSE BRIGHTFIELD page by
clicking MODES and then CONFIGURATION.)
- Once the
aperture has been selected, center it using both the smaller knurled
knob in the series on the aperture assembly and the small knurled knob
to the right. Remember not to manipulate the smallest, innermost knob,
which is not knurled and will unscrew the aperture rod.
- Press D to exit
diffraction mode.
|
| Objective
Lens Astigmatism Correction: |
- Select
an area that may be imaged at high magnification without harming any desirable
portions of the specimen. Increase the magnification to 175,000x or higher and
adjust the illumination so that the substructure or background grain of the
specimen may be observed easily. The INTENSITY will have to be modified and the
beam will have to be recentered using the SHIFT X/Y knobs (Deflectors) as the
magnification is increased. This latter function may alternatively be
controlled using the RST button, on the panel to the left of the column.<
- Set the focus step size to 2 and
obtain a slightly underfocused image for maximum contrast.
- Press the STIG button to open the STIGMATOR CONTROL page. If it is not already highlighted,
press the OBJ key on this page.
- Use the MULTIFUNCTION X/Y knobs, one at a time, to obtain the sharpest possible image of the
grain substructure.
- Confirm that any astigmatism has been corrected by varying the focus (back and forth through
focus, from underfocus to overfocus) and watching to see if a "streaking" pattern emerges and changes
direction between under- and overfocus. If the astigmatism has been corrected, the specimen will vary only
in focus, with no pattern evident. Repeat steps 4 and 5 until this state is
achieved.
- Press the STIG button again to return to the TEM
LOW DOSE BRIGHTFIELD page.
|
Sponsors: National Science
Foundation BIR-9317781
|